Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Data above occulter mask all black #250

Open
ulisapg opened this issue Feb 17, 2024 · 2 comments
Open

Data above occulter mask all black #250

ulisapg opened this issue Feb 17, 2024 · 2 comments
Assignees
Labels
bug fix of a problem in the existing code
Milestone

Comments

@ulisapg
Copy link
Collaborator

ulisapg commented Feb 17, 2024

Only occulter mask visible in frames on the web:

Dates Times
20220202 195044
20220310 195208, 200822 (both doppler)
20220311 193315
20220313 194616, in doppler 200551, 203925
20220611 194735
20220612 015723, 011047 (doppler)
20220902 174632, 174904, 182443, 182715, 190516

@ulisapg ulisapg added the bug fix of a problem in the existing code label Feb 17, 2024
@mgalloy mgalloy self-assigned this Feb 17, 2024
@mgalloy mgalloy added this to the 1.1 reprocessing milestone Feb 17, 2024
@bberkeyU
Copy link
Contributor

The 20220202 195044 It looks like the Lyot filter temperature was about 1 degree low, and for some reason (I forgot what we were testing), temperature compensation was turned off. We should remove 20220202.195044 from future processing. This may need to be epoch-based because it may be hard to come up with a criteria that would cover this issue.

The Lyot filter is also too cool for the next 3 files, but these were rejected for GBU =16 identical temps. We got lucky here since the data would have been poor quality. But there is only one pair of files with identical temps, so I dont understand why 3 files got tagged with Reason 16. I believe 20220202.195251.ucomp.1074.l1.p5.fts and 20220202.195705.ucomp.1074.l1.p5.fts should pass the GBU with reason 0. 20220202.195457.ucomp.1074.l1.p5.fts should still have reason 16 since it matches 20220202.195457

The next group appears to have bad frames during wave data; the new pipeline code should catch the missing states and process them to L1.
20220310 195208, 200822 (both doppler)
20220311 193315
20220313 194616, in doppler 200551, 203925
20220611 194735
20220612 015723, 011047 (doppler)

In 20220902, Lyot filter temps start the morning very low and only appear to come up to temperature about 17:50UT in the unfiltered temps. Unfortunately, the filtered LCVR temps stay low until about 20UT, so most/all of the morning data is probably suspect.

@bberkeyU
Copy link
Contributor

My comment about reason 16, seen above, was based on an incorrect understanding of how we calculated identical temps. I remember it as identical temps between two fits files and not multipole RTD probes measuring the same value.

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment
Labels
bug fix of a problem in the existing code
Projects
None yet
Development

No branches or pull requests

3 participants